-
其它传感器
您现在所在的位置:首页 > 其它传感器
4CFC-一氧化碳传感器
类 别: 其它传感器
描 述: Operating Principle 3-electrode electrochemical Measurement Range 0-1000 ppm CO Maximum Overload 2000 ppm CO
13500022617
Technical Specifications:
| MEASUREMENT | |
| Operating Principle | 3-electrode electrochemical |
| Measurement Range | 0-1000 ppm CO |
| Maximum Overload | 2000 ppm CO |
| Filter | To remove acid gases andhydrocarbons |
| Sensitivity* | 0.7 + 0.15 μA/ppm |
| Response Time (T90)* | 20 Seconds at 20℃ |
| Baseline Offset (clean air)* | <+2 ppm equivalent |
| Zero Shift (-40℃ to +50℃)* | < +12 ppm equivalent |
| Repeatability | <+3% |
| Linearity | Linear<±5% |
| ELECTRICAL | |
| Recommended Load Resistor | 5Ω |
| Bias Voltage | Not required |
| MECHANICAL | |
| Housing Material | Noryl 110 |
| Weight | 5 g(approx.) |
| Orientation | Any |
| ENVIRONMENTAL | |
| Typical Applications | Portable life safety |
| Operating Temperature Range: | |
| Continuous | -20'℃ to +40℃ |
| Intermittent | -40℃to +55℃ |
| Operating Pressure Range | 1 atm+ 10% |
| Operating Humidity Range | |
| Continuous | 15% to 90% RH non-condensing |
| INTRINSIC SAFETY DATA | |
| Maximum at 2000 ppm | 0.2 mA |
| Maximum o/c Voltage | 1.3 V |
| Maximum s/c Current | <1.0 A |
| LIFETIME | |
| Long Term Output Drift | < 5% per annum |
| Recommended Storage Temp | 10℃ to +30℃ |
| Expected Operating Life | 2 years in air |
| Storage Life | 6 months in original packaging |
| Standard Warranty | 18 months from date of despatch |
Specifications are valid at 20°c, 50% RH and 1013 mBar, using City Technology recommendedcircuitry. Performance characteristics outline the performance of sensors supplied within thefirst 3 months. Qutput signal can drift below the lower limit over time.